| Insulation
resistance |
| Test
voltage |
100V |
250V |
500V |
1000V |
Measuring
range
(Auto range) |
4.000/40.00/
200.0M次 |
4.000/40.00/400.0/2000M次 |
| First
Effective Measuring range |
0.1~20M次 |
0.1~40M次 |
0.1~200M次 |
0.1~1000M次 |
| Mid-scale
value |
5M次 |
50M次 |
| Accuracy |
|㊣(2%rdg+6dgt) |
| Second
effective measuring range lower |
0~0.099M次 |
| Second
effective measuring range upper |
20.01~
200.0M次 |
40.01~
2000M次 |
200.1~
2000M次 |
1001~
2000M次 |
| Accuracy |
|㊣(5%rdg+6dgt) |
| Rated
current |
DC
1~1.2mA |
| Output
short circuit current |
1.5mA
max |
| 次/Continuity |
| Auto
range |
40.00/400.0次 |
| Accuracy |
|㊣(2%rdg+8dgt) |
| Output
Voltage on open circuit |
5V㊣20% |
| Output
short circuit current |
DC 220㊣20mA |
| Fuse |
Quick
acting ceramic fuse 0.5A/600V (Ø6.35℅32mm) |
| AC Voltage |
| Range |
AC
20~600V (50/60Hz)
DC -20~-600V/+20~+600V |
| Accuracy |
|㊣(3%rdg+6dgt) |
| General |
| Other
function |
Data
hold, Zero 次 ADJ., Auto discharge, Bar graph
display, Memory, Comparator, Live circuit warning,
Auto power off,
Battery alarm, Backlight, Safety lock, Remote
measurement,
Measurement button lock(Continuous measurement) |
| Safety
standard |
IEC/EN
61010-1 : 2001 CAT.III 600V IEC/EN 61557-1,2,4 |
| EMC
standard |
IEC/EN
61326-1 (CE marking) |
| IP
rating |
IP40 |
| Dimensions
/ Weight |
105(L)
℅ 158(W) ℅ 70(D)mm / 600g approx. |
| Power
source |
R6P℅6
or LR6℅6 |
| Accessories |
7103
(Test lead with remote control switch),
7131 (Safety crocodile clip), 7161
(Test bar),
8017
(Extension prod), 9120 (Cord case), 9121 (Shoulder
strap), R6P (AA)℅6, Instruction Manual |
| Optional |
7115
(Extension Probe), 8016
(Pickel Type Prod),
8923 (Fuse [F500mA/600V]), 9089 (Carrying Case) |